The MuCAP magnet setup consists of the following items:
- the magnet coils having 2 attached temperature switches
- the magnet switch box which switches off the magnet supply current at temperatures
above 35 degree C
- the power supply
- the water pump for the cooling circuit (located behind the concret door)
2) turn on the magnet switch box
- switch the on/off button to on
- press the RESET / Magnet ON button - you will hear a click of the relay
3) turn on the magnet power supply
- lift the black handle-switch on the left side of the front panel
There are 2 turning knobs: the left constrains the voltage, the rigth
constrains the current
- turn the left knob (voltage) to allow maximum voltage
- turn slowly the right knob (current), and increase the current to its maximum
(~200 A on the display)
200 A current in the coil should give about 24 degrees C on each coil.
- press the on switch on the thermoelement display.
- press T1 to see the temperature of the upper coil
- press T2 to see the temperature of the lower coil
- look at the front panel display of the power supply
- use a voltmeter and check the voltage between the 2 copper tubes which
are connected to the
power supply. The voltage at 200 A should be 9.40 V.
- use the Gauss-meter which is in the area
1) turn the right know (current) on the front panel of the power supply
back till the current
monitor displays = 0
2) flip down the black flip switch of the power supply
3) switch off the magnet switch box
4) turn off the water pump
1) Power supply does not have power:
- check if switch box is on and
press reset/magnet on
- check if the 2 temperature switches
are plugged into the switch box
(each
has a red and white plug on a yellow cable)
- check if power cord is plugged
into switch box extension cord
- check if switch box power plug
1 is plugged into 3 phase power
- check if switch box power plug
2 is plugged into 110 Volt transformer
- magnet is warmer than 35 degree
C -> check water pump
if magnet
was too hot -> check water pump and wait till temperature switches cool
off < 35
2) Power supply has power, but current does not turn on
- check if voltage (right knob
is turned to maximum)
- check if all power connections
on the magnet and on the back of the power supply are ok
3) Pump does not work anymore
- check if power cable is plugged
in
- call PSI cooling group Hr. Strittmatter
tel: 3035
4) Water is leaking out of the conectors on the hose-magnet connection
- switch off the pump
then
try to tighten the connection by using new teflon tape
if
this is not successful, try to use a new connector, which is in one of the
magnet supply boxes in the area
There are 4 red hoses, 2 connect on one side to the
outlet (VL = Vorlauf) of the pump system,
2 connect to the return line (RL = Ruecklauf) of the
pump system
Every coil connects to one VL and one RL hose.
The pressure on the VL line can be set with the
2 outlet valves and should be around 3-4 bar,
Electrical connections:
see the electric circuit diagram